Comparison of corneal thickness measurements using pentacam, orbscan II, and A-ultrasonic pachymetry in myopia

LEI Yu-lin1, ZHENG Xiu-yun1, DANG Guang-fu2, LIU Su-mei1   

  1. 1. Department of Ophthalmology, Jinan Mingshui Eye Hospital, Zhangqiu 250200, Shandong, China;
    2. Department of Ophthalmology, Shandong Qianfoshan Hospital, Jinan 250014,  China
  • Received:2010-06-25 Revised:2010-12-11 Online:2011-04-16 Published:2011-04-16

摘要:

目的    比较Pentacam、OrbscanⅡ和A超三种角膜测厚仪检测近视眼中央角膜厚度的差异。  方法    分别用Pentacam眼前节分析仪、OrbscanⅡ角膜地形图仪及A超对角膜厚度进行检测,并将测量值进行统计学分析。结果     Pentacam测得角膜厚度为(536.01±32.33 )μm,OrbscanⅡ测得近视眼角膜厚度为(539.21±37.45)μm, A超测得角膜厚度为(535.39±31.46)μm。OrbscanⅡ测量值与Pentacam和A超测量值差异均有统计学意义(P<0.001),Pentacam与A超测量值间的比较差异无统计学意义(P=0.095)。三种测厚结果显示高度正相关,Orbscan-Pentacam、Orbscan-A超、Pentacam-A超的相关系数r分别为0.917、0.906、0.965,回归方程如下:Orbscan(μm) = -30.699 + [1.0622 × Pentacam(μm)], Orbscan(μm) =-38.599 + [1.0792 × A超(μm)], Pentacam(μm) = 5.424 + [0.9919 × A超(μm)]。结论     在检测中央角膜厚度方面,OrbscanⅡ与Pentacam及A超不可替代使用,Pentacam与A超可相互替代使用。 

关键词: 角膜厚度;近视眼;Pentacam;OrbscanⅡ;A型超声

Abstract:

Objective    To compare the central corneal thickness measurements using pentacam,orbscanⅡand A-ultrasonic pachymeter in myopic eyes. Methods     Central corneal thickness was measured and analyzed with pentacam, orbscan II, and A-ultrasonics.Results     The mean central corneal thickness measured by pentacam,orbscanⅡ,and A-ultrasonics were (536.01±32.33) μm, (539.21±37.45)μm,and(535.39±31.46)μm,respectively. Significant difference was found between orbscanⅡ and pentacam or A-ultrasonic (P<0.001). No significant difference was found between pentacam and Aultrasonic(P=0.095). Positive relationships were found in the results of corneal thickness among the 3 methods.The correlation coefficient r of (orbscanpentacam), (orbscan-A-ultrasonic), (pentacam-A-ultrasonic) were correpondinging 0.917,0.906,and  0.965. The regression equations were “orbscan(μm) = -30.699 +[1.0622 × pentacam(μm)]”, “orbscan(μm) =-38.599 + [1.0792 × A-ultrasonic (μm)]”, and  “pentacam(μm) = 5.424 + [0.9919 × A-ultrasonic (μm)]”. Conclusion    For central corneal thickness, pentacam is superior to A-ultrasonic, but  orbscan /pentacam and orbscan/A-ultrasonic are not substitutable for each other.

Key words: Corneal thickness; Myopia;  Pentacam; OrbscanⅡ; A-ultrasonic